Abstract
AbstractA live retention database for compound identification in isothermal and any step temperature programmed capillary gas chromatography has been developed. The database utilizes the Kovats retention indices of compounds on a given stationary phase and the retention time of n‐alkanes measured at isothermal conditions on the column to be used, together with the programming parameters. Identification is performed by search operation that compares the calculated results with the retention values of unknown peaks. Cross‐reference of the search results of different operating conditions is performed automatically by the database in order to increase the reliability of the identification. The error of the database conversion is ≤ ± 0.5 index unit, or ≤ ± 1% on retention time. This paper describes the principle and the structure of the database in detail. The experimental results for different calsses of compounds tested at divers operating conditions will be presented in Part II.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.