Abstract
A live motile sperm sorting device (LensHooke® CA0) developed to prevent the deleterious effects of centrifugation was evaluated comparatively with conventional density-gradient centrifugation (DGC) and microfluidic-based device (Zymot) in sperm selection. Semen samples from 239 men were collected. CA0 under different incubation intervals (5, 10, 30, and 60min) and temperatures (20, 25, and 37℃) was conducted. The sperm quality in CA0-, DGC-, and Zymot-processed samples was then comparatively evaluated. Semen parameters included concentration, motility, morphology, motion kinematics, DNA fragmentation index (DFI), and the rate of acrosome-reacted sperm (AR). Total motility and motile sperm concentration increased in a time- and temperature-dependent manner and the total motility peaked for 30min at 37℃. In paired analysis, CA0 showed significantly higher total motility (94.0%), progressive motility (90.8%), rapid progressive motility (83.6%), normal morphology (10.3%), and lower DFI (2.4%) and AR (4.7%) than the other two methods in normozoospermic samples (all p < 0.05). For non-normozoospermic samples, CA0 had significantly better results than the other two methods (total motility 89.2%, progressive motility 80.4%, rapid progressive motility 74.2%, normal morphology 8.5%, DFI 4.0%, and AR 4.0%; all p < 0.05). CA0 yielded spermatozoa with enhanced sperm fertilization potentials; DFI was minimized in samples processed by CA0. CA0 was effective for both normal and abnormal semen samples due to its consistent selection efficiency.