Abstract

Recent advances in the computation of shock wave front propagation [P. T. Christopher and K. J. Parker, J. Acoust. Soc. Am. 90, 488–499 (1991)] have allowed the prediction of spark-gap lithotripter waveforms both on and off-axis for a variety of geometries and conditions. Since in both urinary and biliary lithotripsy, shock pulses must propagate through layered human tissues (e.g., fat, muscle, and other soft tissues) it is especially important to understand how waveforms evolve in layered media. Comparisons of computational predictions and experimental results are presented for a laboratory lithotripter of our own construction whose axis of symmetry is vertical. Measurements of waveforms were made using a novel PVDF membrane hydrophone [E. C. Everbach, J. Acoust. Soc. Am. Suppl. 1 87, S128 (1990)] immersed in oil–water layers horizontally stratified by buoyancy forces. The comparisons show sufficient agreement to allow the possibility of optimizing lithotripter reflector geometries so that waveforms could be tailored to accommodate tissue configurations of individual patients.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.