Abstract

In this work, we report on the implementation of a closed-loop liquid-loaded resonant mass sensor as a test vehicle for the realization of highly sensitive miniaturized biomarker assays. Using thin-film piezoelectric-on-silicon (TPoS) contour-mode disc resonators that are specifically designed and optimized for liquid-phase operation, a liquid-loaded low power (~1.6 mW) oscillator at ~17 MHz is demonstrated with a single transistor. By proper placement of the electrodes and the supporting tethers, a high liquid-loaded quality factor (~380) and a low insertion loss (~19 dB) are concurrently achieved which enable the accurate frequency tracking of the potential mass microbalance. The frequency shifts due to the liquid-loading and the preliminary characterized frequency stability of the oscillator imply a minimum mass detection resolution in the picograms range. [2020-0177]