Abstract
A liquid chromatography–electrospray ionization tandem mass spectrometry (HPLC–ESI-MS/MS) method for the simultaneous quantitation of artemether and lumefantrine in human plasma was developed and validated. Artesunate was used as an internal standard (IS). The analytes were extracted by a protein precipitation procedure and separated on a reversed-phase Zorbax SB-Ciano column with a mobile phase composed of methanol and 10 mM aqueous ammonium acetate containing 0.2% (v/v) acetic acid and 0.1% (v/v) formic acid. Multiple reaction monitoring was performed using the transitions m/ z 316 → m/ z 267, m/ z 530 → m/ z 348 and m/ z 402 → m/ z 267 to quantify artemether, lumefantrine and artesunate, respectively. Calibration curves were constructed over the range of 10–1000 ng/mL for artemether and 10–18,000 ng/mL for lumefantrine. The lower limit of quantitation was 10 ng/mL for both drugs. The mean R.S.D. values for the intra-run precision were 2.6% and 3.0% and for the inter-run precision were 3.6% and 4.6% for artemether and lumefantrine, respectively. The mean accuracy values were 102.0% and 101.2% for artemether and lumefantrine, respectively. No matrix effect was detected in the samples. The validated method was successfully applied to determine the plasma concentrations of artemether and lumefantrine in healthy volunteers, in a one-dose pharmacokinetic study, over the course of 11 days.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.