Abstract

Galunisertib is an anti-cancer drug currently evaluated in phase I and II clinical trials. This study describes the development and validation of a bioanalytical assay to quantify galunisertib in human plasma using HPLC-MS/MS. Stable isotope labelled galunisertib was added as internal standard and the analyte and internal standard were extracted from the matrix by protein precipitation using acetonitrile-methanol (50:50, v/v). Final extracts were injected onto a C18 column, gradient elution was applied for chromatographic separation and detection was performed using a triple quadrupole mass spectrometer operating in the positive ion mode. The assay was linear over the range 0.05–10 ng/mL, with acceptable accuracy (bias ranging from −6.1 to 3.1%) and precision (below 5.7% C.V.) values. The applicability of the assay was demonstrated in a pharmacokinetic experiment in mice.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.