Abstract
A quantitative bioanalytical liquid chromatography–tandem mass spectrometric assay for the tyrosine kinase inhibitor pazopanib was developed and validated. Plasma samples were pre-treated using protein precipitation with acetonitrile containing pazopanib-d4 as internal standard. The extract was injected into the chromatographic system after dilution with water (1:9, v/v). The system consisted of a sub-2μm particle, trifunctional bonded octadecyl silica column with isocratic elution using 0.005% (v/v) of formic acid in a mixture of water (76%, v/v) and acetonitrile (24%, v/v). The analyte was quantified using the selected reaction monitoring mode of a triple quadrupole mass spectrometer with a heated electrospray interface. The assay was validated in a 0.1–100μg/ml calibration range. Within day precisions were 3.6–5.2%, between day precisions 4.0–8.3% and accuracies between 106% and 113% for the whole calibration range. The drug was sufficiently stable under all relevant analytical conditions. The assay has successfully been used to assess drug levels for therapeutic drug monitoring in patients treated with pazopanib.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.