Abstract
A quantitative bioanalytical liquid chromatography-tandem mass spectrometric (LC-MS/MS) assay for the mutated BRAF inhibitor dabrafenib was developed and validated. Plasma samples were pre-treated using protein precipitation with acetonitrile containing PLX4720 as internal standard. The extract was directly injected into the reversed-phase chromatographic system after dilution with water. The eluate was transferred into the electrospray interface with positive ionization and the analyte was detected in the selected reaction monitoring mode of a triple quadrupole mass spectrometer. The assay was validated in a 2-2000 ng/ml calibration range with r(2)=0.993±0.002 for linear regression with quadratic weighting (n=5). Within day precisions (n=6) were 3.3-5.2%, between day (3 days; n=18) precisions 4.7-8.2%. Accuracies were between 95-104% for the whole calibration range. The drug was sufficiently stable under all relevant analytical conditions. Finally, the assay was successfully used to determine drug pharmacokinetics in mice.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.