Abstract

Microcystins (MCs) are toxins produced by cyanobacteria which are common organisms in the phytoplankton of eutrophic lakes, rivers and freshwater reservoirs. In the present work, a novel method of liquid chromatography–electrospray ion trap tandem mass spectrometry (LC/ESI/Ion trap–MS/MS), operated in the negative ionization mode, was developed for the analysis of these cyanotoxins. The method was applied to determine the amounts of total microcystins-LR, -YR and -RR in two water reservoirs in Southern Portugal, namely Alqueva and Beliche. A total of 30 water samples were analysed along 2011. Solid phase extraction (SPE) was used for sample cleaning-up and analyte enrichment. The extracted toxins were separated on a C18 column with a gradient of acetonitrile/water with 0.1% formic acid. Detection of microcystins was carried out using multiple reaction monitoring (MRM) in the negative polarity mode, as this method gave a higher selectivity. The MC-RR, YR and LR quantification limits were 17.9, 31.7 and 15.8 ng/L, respectively; quite below the limits recommended by WHO guidelines for drinking water (1 μg/L). Total MC highest concentrations were found in the warm months of June, July and September in Alqueva sampling sites, with concentrations of MC LR and RR ranging 17–344 and 25–212 ng/L, respectively, showing comparable results for MC-RR and LR and slightly lower concentration of MC-YR. Detected values for Beliche reservoir were below quantification limits.