Abstract
A new simple and rapid liquid chromatographic–mass spectrometric technique was designed for the determination of nine benzodiazepines in plasma and oral fluid. Benzodiazepines were extracted from alkalinised spiked and clinical plasma and oral fluid samples using a single step, liquid–liquid extraction procedure with diethyl ether. The chromatographic separation was performed with a Xterra ® RP 18, 5 μm (150 × 2.1 mm i.d.) reversed-phase column using deuterated analogues of the analytes as internal standard. The recovery ranged from 70.3 to 86.9% for plasma and 63.9 to 77.2% for oral fluid. The limits of detection ranged from 0.5 to 1 ng/ml in plasma and 0.1 to 0.2 ng/ml for oral fluid. The method was validated for all the compounds, including linearity and the main precision parameters. The procedure, showed to be sensitive and specific, was applied to real plasma and oral fluid samples. The method is especially useful to analyse saliva samples from drivers undergoing roadside drug controls.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.