Abstract
A robust, accurate and sensitive high-performance liquid chromatographic method for the determination of rosiglitazone ( I) in human plasma has been developed. Pioglitazone ( II) was used as internal standard. Both I and II are extracted from plasma using a liquid–liquid extraction procedure. Isocratic separation of I and II is carried out using a reversed-phase Zorbax SB C 18, 15-cm column with mobile phase consisting of methanol and a mixed phosphate buffer (10 m M monobasic sodium phosphate and dibasic sodium phosphate, pH adjusted to 2.6 with ortho-phosphoric acid) in the ratio 30:70 (v/v) and quantified by UV detection at 245 nm. Linearity was established over the range 5–1250 ng/ml using 1 ml human plasma. The method is specific, the endogenous components in plasma do not interfere with I and II. C.V. (%) of intra-day samples is less than 5.0% at four concentrations tested namely 5, 10, 500 and 1000 ng/ml. Similarly, over the same nominal concentrations, the precision of inter-day (5 days) samples also results in C.V. (%) less than 5.0%. The recoveries of I and II from human plasma were about 79 and 60%, respectively. This method can be used for routine clinical monitoring of I.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.