Abstract

A high-performance liquid-chromatographic (HPLC) assay with UV detection has been developed for the quantitative determination of erlotinib (OSI-774) in human plasma. Quantitative extraction was achieved by a single-solvent extraction involving a mixture of acetonitrile and n-butyl chloride (1:4, v/v). Erlotinib and the internal standard hydrochloride salt (OSI-597) were separated on a column packed with Nova-Pak C18 material and a mobile phase composed of acetonitrile and water, pH 2.0 (60:40, v/v). The column effluent was monitored with dual UV detection at wavelengths of 348 nm (erlotinib) and 383 nm (OSI-597). The calibration graph was linear in the range of 100–4500 ng/ml, with values for accuracy and precision ranging from 87.9 to 96.2% and 2.13 to 5.10%, respectively, for three different sets of quality control samples. The developed method was successfully applied to study the pharmacokinetics of erlotinib in a cancer patient at the recommended daily dose of 150 mg.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.