Abstract

Abstract A liquid chromatographic method for the determination of amiodarone hydrochloride and 10 related compounds in drug raw material and for assay of drug in tablets was developed. The method specifies a 3 jxm Hypersil nitrile column (150 × 4.6 mm), a mobile phase of 1 + 1 acetonitrile–ammonium acetate buffer (0.1 M adjusted to pH 6.0 with 0.1 M acetic acid), a flow rate of 1 mL/min, and detection at 240 nm. The lower limit of quantitation of the related compounds is 0.02% or less. Drug contents in 2 raw material samples were 100.1 and 99.9% and ranged from 98.2 to 99.4% in 3 tablet formulations. Impurity levels in 2 samples of raw material from different manufacturers were ca 0.4%. The presence of 3 of the known related compounds in these samples was confirmed by liquid chromatographymass spectrometry. The method applied to raw materials was evaluated by a second laboratory and found to be satisfactory.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.