Abstract
BackgroundCeramic-on-ceramic (CoC) bearings for total hip arthroplasty (THA) have been offering very favorable results and survivorship since their introduction. In order to increase range of movement (ROM) and decrease dislocation rates, some manufacturers have introduced larger diameter head (LDH) CoC bearings. This has been achieved with the use of preassembled cup designs, in which the ceramic liner is already fitted into the metal backing and implanted as a monoblock component by the surgeon. In this report we present data from a series of 5 patients with ceramic liner dissociation from a monoblock cup.Case presentationAll cases were overweight men with acetabular components of 56 or 58 mm. After a mean of 5.5 (range, 3.5-6.7) years, all patients reported sudden pain and audible noise when performing activities of daily living. Liner displacement was suspected on plain radiographs and confirmed by Ct-scan. Pneumarthrosis was present in all cases. Taper modular junction wear and corrosion signs were observed in the four revised patients.ConclusionAlthough one of our case is still treated conservatively, implant revision is probably inevitable. Further LDH CoC implant design should take in consideration this potential complication by avoiding bearing diameters over 40mm and/or improving locking mechanism or by providing a real monoblock acetabular implant.
Highlights
Ceramic-on-ceramic (CoC) bearings for total hip arthroplasty (THA) have been offering very favorable results and survivorship since their introduction
Further larger diameter head (LDH) CoC implant design should take in consideration this potential complication by avoiding bearing diameters over 40mm and/or improving locking mechanism or by providing a real monoblock acetabular implant
Modern ceramic has enabled minimization of clinical fractures [3] and introduction of larger diameter head (LDH) CoC bearings in order to increase range of movement (ROM) and decrease dislocation rates, while avoiding the associated problems seen in metal-on-metal (MoM) bearings
Summary
One of our case is still treated conservatively, implant revision is probably inevitable.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.