Abstract

This paper presents a novel technique developed for linearizing the energy spectra of radiation detectors in commercial radioisotope identification devices. Based on few spectrum measurements with standard radio-nuclide sources, this method allows generation of individual nonlinear calibration functions at minimum expense in the routine instrument setup. Instead of fitting peak positions, the measured raw data are compared with simulated spectrum templates, and local gain factors providing the best correspondence are taken as reference points for the calibration function. This approach avoids the problem of fitting multiple peaks with intensity ratios influenced by absorbing layers and assures an accuracy of 1% in the energy range of 30 keV to 3 MeV.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.