Abstract
The accuracy of the dose monitor system of a medical electron accelerator is of crucial importance for the precise delivery of the prescribed dose. Special attention is given to the case of low monitor units (MU). The long-term stability of the dose monitor system is analyzed. Photon and electron fields with monitor units between 1 and 500 MU were measured with field sizes of 10 x 10 cm(2) and 20 x 20 cm(2) at source-skin distances (SSDs) of 100 cm and 95 cm, respectively. The relative dose per MU [see text] was determined for an Elekta linear accelerator. The relative dose per MU was measured to be constant within about 1% down to fields with only 2 MU. The deviation remained below 2% down to 1 MU which is the minimum setting of the accelerator. The long-term stability is better than 0.2% per month. No therapeutically relevant deviation of the prescribed dose could be detected for the Elekta linear accelerator. This is a prerequisite for the implementation of intensity-modulated radiotherapy (IMRT) techniques.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.