Abstract
The half-life of 99 mTc was measured at the JRC using the ionisation chamber 'IC1' (type Centronic IG12). The result, T1/2(99 mTc) = 6.00660 (18) h, is in good agreement with literature data. One experiment was performed in IC1's default set-up with the ionisation current being integrated over an air capacitor and read out as a voltage increase over time. This ensured excellent linearity and precision throughout the dynamic range, but the maximum current was limited to 2 nA. In a second test, the current was directly read out with a Keithley 6517 A electrometer. Applying correction factors for the automatic range switching of the electrometer, an acceptable linearity was demonstrated over a range of 12 half-life periods starting at 20 nA. Range switching and autocorrelation of the current readout increase the systematic and random error propagation factors. Piecewise fitting of the decay curve over periods of 6 h yields the same 99 mTc half-life value within 0.04% (0.0025 h) standard deviation over an activity range spanning at least 10 half-life periods (3 orders of magnitude).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.