Abstract

Background Radiofrequency (RF)-assisted devices are widely used for hemostasis during liver resection. This study compared the use of dual switching (DS) versus single switching (SS) control modes for RF-based liver resections in a pig model. Methods The RF-based system comprised a 200-W generator and three electrodes with 4-cm tips arranged in a linear configuration using an adaptor. Eight Lanyu pigs were used to assess ablation outcomes with electrode spacing of 2 or 3 cm, and ablation durations of 1.5, 2 or 3 min. All combinations were tested in DS and SS modes. Procedures were performed on left lateral, caudal and right anterior liver lobes, and after which transections were performed using a scalpel. Blood loss, complete ablation rate and ablation speed were compared. Results DS mode was shown to induce significantly less blood loss than SS mode when the electrode spacing was set at 2 cm and the ablation duration was 2 min or 3 min (p=.010 and .012, respectively). Extended ablation duration and narrow electrode spacing tended to induce less blood loss, regardless of operating mode. Bloodless resection was achieved using DS mode with electrode spacing of 2 cm and ablation duration of 2–3 min. The highest rate of complete ablation (11.3 cm2/min) was achieved using DS mode with electrode spacing of 2 cm and ablation duration of 1.5 min. Conclusion RF-based hepatic resection using DS mode is safe and feasible, resulting in less blood loss than SS mode with a higher rate of complete ablation (i.e., superior ablation efficiency).