Abstract

To evaluate whether the measurement accuracy of dental CT images is affected by the mandibular positioning and the pitch factor at CT scanning. Three dry mandibles were scanned by 64-slice multidetector row CT (MDCT) and micro-CT. For MDCT scanning, 7 different mandibular positioning and 3 different pitch factors were applied. After reformatting dental CT images, the bone height was measured on these images. It was also measured on the corresponding micro-CT image, which was defined as the actual value. The difference of the measurement values between these 2 was defined as the measurement error. There was no significant difference in the measurement errors due to either the mandibular positioning or the pitch factor. The measurement accuracy of dental CT images obtained was not affected by either mandibular positioning or pitch factor at CT scanning.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.