Abstract
Linear and volumetric wear was measured in 33 tibial polyethylene inserts from three different cruciate-retaining knee systems retrieved at the time of revision surgery. Wear patterns also were evaluated and classified. Eccentric and asymmetric wear patterns were seen in 78% of inserts with flat articulating geometry versus 12% in inserts with curved anteroposterior geometry. The mean linear wear rate was .35 mm/year (range, .05-1.68 mm/year) and the mean volumetric wear rate was 794 mm3/year (range, 24-4088 mm3/year). Linear and volumetric wear rates showed a negative correlation with the length of implantation. Linear wear rates also showed a negative correlation with patient weight.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.