Abstract
An expert system for control of a rapid scanning spectrometer to perform atomic emission spectrometry has been developed using the PROLOG language. The expert system can be “taught” the elemental composition of a given sample type. It uses this information (when available) to generate a line-search strategy which minimises the number of lines that need to be measured. The rules used by the system can be observed during system operation to allow a trace of the logic. Reading of the initial database takes no more than 6 s and then decisions are made in less than 1 s per line.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.