Abstract
The total intrinsic variability of pulsed Doppler spectra of mitral blood flow was evaluated in this study. Doppler examinations were performed in 10 normal volunteers (7 men, 3 women), ages, 29 to 57 years (mean 41) with an interval period of 3 months. The recordings were made in the apical 4-chamber view, using a 2.25-MHz transducer. The sample was placed approximately 1 cm proximal (left atrium) and 1 cm distal (left ventricle) from the middle of the anulus of the mitral valve at the onset of diastole. Processing of the raw Doppler spectra was done in a way that avoided subjective manipulation by the investigators. Mitral peak velocities, accelerations and spectral widths were measured in early diastole and late diastole. From the parameters under investigation, maximal velocities showed the best reproducibility with a range of coefficients of variation of 9 to 13%. Peak acceleration in early diastole and spectral widths showed rather large coefficients of variation, ranging from 19 to 30 and 14 to 24%, respectively. Mitral pulsed Doppler spectra in the left atrium were more reproducible than in the left ventricle.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.