Abstract

Limits of performance for measurement of blood haemoglobin concentration, number concentration of erythrocytes, and erythrocyte volume fraction are presented. The limits of performance are based on actual performance by the participants of the Finnish external quality control programme in haematology. In evaluation of a laboratory's performance in measuring blood haemoglobin (Fe) concentration, all values within +/- 0.20 mmol/l of the corrected mean are considered to represent 'good' performance; those between +/- 0.20 and +/-0.40 mmol/l represent 'acceptable' performance and values outside +/- 0.40 mmol/l are 'not acceptable'. Similarly, in measuring number concentration of erythrocytes the limits of +/- 0.2 x 10(12)1(-1) and +/- 0.4 x 10(12)1(-1) and in measuring erythrocyte volume fraction the limits of +/- 0.01 and +/- 0.02 were established. The limits of performance proved to be readily attainable by the survey participants.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.