Abstract

An X-ray tube with a Mo target and Zr filter, operated at 45 kV/20 mA, was used to excite samples (5 microL deposited on a quartz support) and the total reflection angle condition was obtained with a double reflector module built with two 10-cm-long 7-mm-thick quartz crystals placed 50 microns apart. A high-resolution spectrometer based on a Si(Li) detector coupled to a multichannel analyzer was used for X-ray detection and the spectra were interpreted with the AXIL software. The system was calibrated with standard chemical solutions containing Cr, Fe, Cu, Zn, and Pb, and Y was used as an internal standard to correct eventual geometric errors and high-voltage instabilities of the X-ray generator. The limits of detection were 19, 9, 5, and 4 ng/mL for Cr, Fe, Cu, and Zn, respectively, analyzed through characteristic K alpha X-rays, and 7 ng/mL for Pb, through L alpha X-rays, considering 50 microL samples deposited and dried on a quartz support, to be excited/detected for 1000 s.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.