Abstract

2D materials are by definition just a few atomic layers thick. They are therefore ideal samples for transmission electron microscopy, in the plan‐view geometry. However, 2D materials are typically placed or grown on substrates, which in some cases requires analysis to be performed on cross sections. In this case focused ion beam preparation is often the technique of choice for producing thin lamellae, but damage to the surface of 2D material during imaging and milling must be mitigated. Herein, it is demonstrated that the typically applied electron beam‐assisted deposition of platinum and carbon prior to milling does not provide sufficient protection, and results in significant damage. Instead, it is found that arc‐evaporated carbon—deposited with a standard carbon coater designed for scanning electron microscopy (SEM) samples—can provide sufficient protection, enabling cross‐sectional analysis without detectable damage to monolayer or bilayer samples subsequently prepared by standard focused ion beam preparation procedures.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.