The susceptibilities of 20 strains of vancomycin-resistant enterococci (VRE) with the vanB genotype obtained by using Vitek GPS-418 cards were compared with those obtained by the broth dilution method of the National Committee for Clinical Laboratory Standards (NCCLS) (approved standard M7-A4) and with those obtained by the agar screen method using bile esculin azide agar containing 6 microgram of vancomycin per ml. Although both the broth dilution and agar screen methods disclosed no discordance, Vitek GPS-418 cards yielded a very major error compared with the results obtained by the reference broth dilution method of the NCCLS. Vitek GPS-418 cards were therefore found to have considerable room for improvement for the accurate detection of vanB VRE strains.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.