Abstract
The effect of keyhole data acquisition on quantitative analysis of dynamic MRI was examined. Experiments were performed retrospectively on raw data obtained from clinical dynamic contrast-enhanced breast imaging procedures. The effects of keyhole phase-encoding acquisition and type of reconstruction algorithm on the accuracy of derived quantitative parameters was assessed. Results indicate that the minimum keyhole size used should be restricted by the approximate minimum size of the expected lesions. Furthermore, reconstruction algorithms that offer improved image resolution do not circumvent this restriction.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.