Abstract

Three popular variations of the iodometric determination of ozone were investigated using a spectrophotometric ozone monitor as the referee method. Determinations carried out with a neutral buffered potassium iodide reagent acidified to pH 2 with sulfuric acid were observed to give values that were 15–30 percent higher than those obtained with the unbuffered potassium iodide and Masschelein modification methods relative to the referee method. Sampling parameters such as flow rate were shown to have a marked effect on the determinations. Because of the technique‐sensitive nature of the determinations, it is recommended that iodometry not be pursued as a standard method for measuring ozone.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.