Abstract

The objective of this study was to determine the validity of estimation of regurgitant volume by visual assessment of color flow Doppler display. An experimental apparatus was designed that is capable of ejecting precise volumes of echogenic material from one chamber to another under continuous color flow Doppler monitoring. The velocity of flow was altered independently by changing either the size of the orifice through which flow occurred or the ejection rate. In this manner the differential effects of volume and velocity on the color flow Doppler image could be examined. The maximum area encompassed by the color flow Doppler pattern for each ejection was planimetered by using commercially available on-line software. In addition the reviewer in each case applied a subjective grade to the appearance of the color flow jet (1+ to 4+). Comparison was then made of the color flow Doppler appearance of equal volumes flowing at different velocities and of different volumes flowing at different velocities. In the initial series a solution of agitated hetasarch was used. When equal volumes were imaged at different velocities the higher-velocity jet appeared larger, both subjectively (3+ vs 1+) and by measuring the area encompassed in the Doppler flow profile (40.3 +/- 1.8 vs 22.0 +/- 1.4 cm2, p = 0.0001). Furthermore, when different volumes were imaged at different velocities, the smaller volume (3 ml vs 6 ml) appeared larger when it was flowing at higher velocity (3+ vs 2+, 40.3 +/- 1.8 vs 32.4 +/- 1.3 cm2, p = 0.0006). These experiments were repeated with blood, confirming the results of the initial study.(ABSTRACT TRUNCATED AT 250 WORDS)

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call