Abstract

A cross-polarization scheme is presented to filter out the excitation light from the emission spectrum of fluorescent dyes using green light emitting diodes as a light source and a linear charge coupled device as an intensity detector. The excitation light was linearly polarized and was then used to illuminate the fluorescent dyes in the microchannels of a capillary electrophoresis microchip. The detector was shielded by the second polarizer, oriented perpendicular to the excitation light. The fluorescent signals from Rhodamine B dyes were measured in a dilution series with resulting emission signals and four different concentrations of fluorescent dyes were detected simultaneously with the same excitation source and detector. A limit-of-detection of 1 μM was demonstrated for Rhodamine B dye under the optimal conditions.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.