Abstract

Lift-off is a challenging issue in alternating current field measurement (ACFM) technique. The spurious signals caused by lift-off variations are similar to the response signals of defects, resulting in defect misjudgment. In this paper, a lift-off point of intersection (LOI) feature is discovered for the first time from the raw signals of ACFM for the elimination of spurious signals caused by lift-off variations. A theoretical model is established to analyze the effect of lift-off variations on characteristic signals of ACFM. A 3D finite element model is built to analyze the effect of the excitation frequency on LOI. The probe and testing system are established, and the validation experiments are carried out. The results show that the traditional Bx signal is susceptible to the variation of lift-off, and the higher the excitation frequency, the closer the LOI is to a point. The novel Bx signal plotted by LOI amplitude has no spurious signals caused by lift-off variations and can also accurately evaluate the crack depth as with traditional Bx signals.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.