Abstract
We report the case of a 65- year old female patient who was urgently admitted to our institution with hemothorax after pacemaker implantation. The bichamber pacemaker had been implanted one week before admission and the active fixation lead caused perforation of the right atrial wall and penetration of the right lung. Transthoracic echocardiography demonstrated a minimal cardiac effusion but chest X-Ray and computed Tomography (CT) visualised the atrial screw helix outside the atrial wall penetrating through the right lung middle lobe causing hemothorax. After median sternotomy and cut-off of the atrial lead screw helix, we sew the perforated area of the right atrium and the penetrated right lung, sucking away the bloody pleural effusion before closing the thorax. Three weeks later, the cut off atrial lead was extracted through the right subclavian vein and a new one was positioned and fixed again. During the hospital admission the patient was well and free of any symptoms and troubles.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.