Abstract

The technique based on quadrupole mass spectrometry of the material ejected by laser-induced ablation – LIA-QMS is proposed as a tool to measure H/D/T content in carbon co-deposits in the first wall and divertor regions of tokamak reactors. Tungsten tiles exposed in Globus-M2 tokamak were taken for validation experiments. Measurement accuracy was determined by comparison with the results obtained by laser-induced desorption – LID-QMS and conventional thermal desorption spectroscopy (TDS). Total amount of deuterium in hydrocarbon deposits was found to be 2.8 × 1017 D/cm2, when measured by – LIA-QMS, 3.0 × 1017 D/cm2 for LID-QMS and 2.9 × 1017 D/cm2 for TDS. The main uncertainty of D surface concentration are caused by inhomogeneity of the deposit thickness over the sample area or by presence of hydrocarbon species in the released gas.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.