Abstract

Four commercially available leukocyte depletion filters were evaluated using a flow cytometric technique to determine the efficiency of each type of filter. The Sepacell R-500A and the PALL RC50 were the most efficient in leukocyte depletion having a mean depletion percentage of 99.3% and 99.5%, respectively. The PALL RC100 which is used for two units also had a 99.3% depletion with the first unit but with the second unit the depletion dropped to 94.2%. The imugard IG-500 had a 97.3% depletion. Red cell recovery ranged from 87.4% for the PALL RC50 to 92.2% for the Sepacell R-500A.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.