Abstract
Background. The bridge to transplantation with pulsatile mechanical assist devices became a standard procedure for patients deteriorating on the waiting list. Recently, continuous flow axial impeller pumps were introduced to clinical application offering new advantages. Methods. From November 1998 till September 2000, 6 male patients (mean age 53 plus or minus 11 years) with end-stage left heart failure were implanted with a DeBakey ventricular assist device (VAD) axial-flow pump for bridge to transplantation. Results. Three patients were successfully transplanted after 74, 115, and 117 days, respectively. Two other patients died after 25 and 133 days. One patient is still on the device after 108 days. Because of modification of the implantation technique after the first 2 patients, mean pump-flow within the first 3 weeks was increased from 4.3 ± 0.6 L/min to 6.7 ± 0.3 L/min. Patients were put on regular bicycle-ergometer training and improved their exercise capacities up to a mean maximum oxygen consumption of 20.2 mL/kg/min. Conclusions. Initial implants of the DeBakey VAD demonstrated support properties comparable to pulsatile pumps but without significant restrictions for extended use.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.