Abstract

Wear is a major contributor to osteolysis and aseptic loosening of total hip replacements (THR). Both alumina (Al(2)O(3)) and cobalt-chrome (CoCr) femoral heads are commonly used. We investigated wear comparing alumina heads to cobalt-chrome heads against conventional cemented polyethylene (PE) cups for up to ten years. Linear wear was measured with radiostereometry (RSA). Our material was derived from two prospective randomised trials that investigated fixation of femoral stems, not wear, and was evaluated retrospectively (Level III). The mean (95% CI) proximal head penetration was 0.96 mm (0.68-1.23) in the cobalt-chrome group and 0.42 mm (0.30-0.53) in the alumina group at ten years (P = 0.001). The mean (95% CI) 3D penetration was 1.07 mm (0.79-1.35) and 0.53 mm (0.38-0.63), respectively, at ten years (P = 0.001). Alumina heads performed better than cobalt-chrome heads in this study after ten-year follow-up.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.