Abstract

Highly crosslinked ultrahigh-molecular-weight polyethylene (XLPE) shows reduced wear in total hip arthroplasty compared to direct compression-molded polyethylene (compPE); however, minimal research evaluating polyethylene damage in XLPE tibial inserts in total knee arthroplasty exists. We evaluated damage and material properties in retrieved XLPE components at midterm (≥ 2.5years) follow-up. We identified 19 XLPE tibial inserts with ≥ 30months in vivo using our institutional review board-approved implant retrieval system; 19 compPE retrieved inserts were matched based on age at index surgery, body mass index, sex, and length of implantation. Articular surface damage was assessed using a subjective grading system. Swell ratio testing and Fourier-transform infrared spectroscopy were used to measure crosslink density (XLPE) and oxidation (XLPE, compPE), respectively, at loaded and unloaded surface and subsurface regions. CompPE inserts had higher overall damage than XLPE inserts, specifically at the post of posterior-stabilized inserts. The XLPE inserts had lower crosslink density at the loaded surface (0.159mol/dm3) than either unloaded region (0.183mol/dm3). CompPE peak oxidation index (OI) was greater than XLPE peak OI in the loaded and unloaded surface regions (1.67 vs. 0.61 and 1.38 vs. 0.46, respectively). Surface damage and oxidation are reduced in XLPE inserts compared to compPE at midterm follow-up. Peak OI greater than 1.0 in the compPE group suggests that mechanical-property degradation had occurred, a likely cause for increased damage. Longer-term retrievals will determine whether these trends continue. Based on midterm results, XLPE shows an advantage over compression molded PE in total knee arthroplasty.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call