Abstract

Previous studies on pulse-height selection for scintillation cameras have been concerned with static imaging or asymmetric window settings around the photo peak. For count densities typical of those encountered in flow studies and for the customary symmetric window it is shown here, both computationally and experimentally, that a window setting in the range of 25-35% provides the maximum lesion detectability rate for lesions of 0.5- to 2.0-cm diameter.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.