Abstract

The bony fixation of reference marker arrays used for computer-assisted navigation during total hip arthroplasty (THA) theoretically involves the risk of fracture, infection, and/or pin loosening. We asked whether intraoperative assessment of leg length (LL) and offset (OS) changes would be accurate using a novel pinless femoral reference system in conjunction with an imageless measurement algorithm based on specific realignment of the relationship between a dynamic femoral and pelvis reference array. LL/OS measurements were recorded during THA in 17 cadaver specimen hips. Preoperatively and postoperatively, specimens were scanned using CT. Linear radiographic LL/OS changes were determined by two investigators using visible fiducial landmarks and image processing software. We found a high correlation of repeated measurements within and between (both 0.95 or greater) the two examiners who did the CT assessments. Pinless LL/OS values showed mean differences less than 1 mm and correlations when compared with CT measurements.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.