Abstract
The inference that mitral valve replacement (MVR) may produce left ventricular outflow tract (LVOT) obstruction has been made, but no comparative hemodynamic studies with various types of prostheses have been done. The purpose of the present study was to compare the gradients created across the LVOT with MVR in young sheep with small left ventricular cavities. Mitral valve replacement was accomplished using cardiopulmonary bypass and hypothermic cardioplegic arrest. Five animals were used for each of the following valves studied: 25-mm Ionescu-Shiley bovine pericardial valve, 25-mm Hancock porcine aortic valve, 2M-6120 28-mm Starr-Edwards ball-valve prosthesis, 25-mm Björk-Shiley 60-degree flat tilting-disc prosthesis, and 25-mm St. Jude Medical hemidisc valve. Gradients across the LVOT were measured after MVR and then during infusion of isoproterenol hydrochloride (0.05 μ/kg/min). Following MVR, only the Starr-Edwards valve produced an LVOT gradient (32 ± 23 mm Hg). Substantial gradients after MVR were seen, however, with isoproterenol administration with the Ionescu-Shiley (47 ± 4 mm Hg), Hancock (13 ± 8 mm Hg), and Starr-Edwards (65 ± 30 mm Hg) valves but not with the low-profile valves (Björk-Shiley and St. Jude Medical). The results of the present study demonstrate that MVR can produce LVOT obstruction. The greatest degree of obstruction was with the high-profile mechanical and bioprosthetic valves.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.