Abstract

BackgroundLead perforation is one of the major complications of pacemaker implantation, but cases of right ventricular (RV) lead perforation through the septum and left ventricle are rarely reported. We described a rare case of left ventricular (LV) free wall perforation by an RV lead and the management of this complication.Case summaryAn 84-year-old man was admitted with a dual-chamber pacemaker due to pacing failure caused by an RV lead fracture. New lead implantation was performed on the next day, but pacing failure occurred again on the second post-operative day (POD). We found the lead perforation on the fluoroscopy during temporary pacemaker insertion. Computed tomography scan and transthoracic echocardiogram showed that the added lead perforated through both the septum and LV free wall. A new lead was inserted on the fourth POD, and an off-pump open chest surgery for extraction of the penetrating lead was performed uneventfully on the 20th POD.DiscussionWe considered that some features of the lead (SelectSecure 3830-69, Medtronic) may be related to this complication, as the lead was very thin, had a non-retractable bare screw and was inserted with a dedicated delivery catheter. We have to be careful when performing implantation of this kind of lead to avoid such a rare complication.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.