Transcatheter aortic valve replacement (TAVR) is an established treatment alternative to surgical valve replacement in high-risk patients with severe symptomatic aortic stenosis. The current guidelines for TAVR are to upsize transcatheter aortic valve (TAV) relative to the native annulus to secure the device and minimize paravalvular leakage. Unlike surgical stented bioprosthetic valves where leaflets are attached to a rigid frame, TAVs must expand to fit within the native annulus. Fully-expanded circular TAVs have consistent leaflet kinematics; however, subtle variations in the degree of stent expansion may affect leaflet coaptation. The objective of this study was to determine the impact of incomplete TAV expansion on leaflet stress and strain distributions. In this study, we developed finite element models of a 23mm homemade TAV expanded to diameters ranging from 18 to 23mm in 1mm increments. Through dynamic finite element simulations, we found that leaflet stress and strain distributions were dependent on the diameter of the inflated TAV. After complete expansion of the TAV to 23mm, high stress and strain regions were observed primarily in the commissures during diastole. However, 2–3mm incomplete TAV stent expansion induced localized high stress regions within the TAV commissures, while 4–5mm incomplete stent expansion induced localized high stress regions within the belly of the TAV leaflets during the diastolic phase of the cardiac cycle. Increased mechanical stress and flexural deformation on TAV leaflets due to incomplete stent expansion may lead to accelerated tissue degeneration and diminished long-term valve durability.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call