Abstract

This study evaluated the radiation dose reduction, operator acceptance, and patient acceptance of two types of leaded thyroid shields designed for use during intraoral dental radiography. Exposure levels were measured with thermoluminescent dosimeters on three groups of 20 patients undergoing complete mouth (20-film) surveys. Skin entrance dose to the thyroid was 20 mR per complete mouth survey without a shield in place, 12 mR per complete mouth survey with the experimental shield in place, and 9 mR with the commercial shield. Patients and radiologic technologist were surveyed to determine patient comfort and operator acceptability. Patient and operator acceptability were higher for the experimental shield than for the commercial shield.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.