Abstract

The purpose was to determine the incidence of lead fracture in patients with DBS over a long period of time. We present a retrospective study of 208 patients who received 387 DBS electrodes. Fourteen patients had sixteen lead fractures (4% of the implanted leads) and two patients suffered from 2 lead fractures. Of all lead fractures, five patients had the connection between the leads and the extension cables located in mastoids region, ten in cervical area and one in thoracic region. The mean distance from the connection between the electrode and the extension cable and the lead fracture was 10.7 mm. The lead fracture is a common, although long-term complication in DBS surgery. In our experience, the most common site of electrode cable breakage is approximately between 9 and 13 mm from the junction between the lead and the extension cable. The most important cause of lead fracture is the rotational movement of the lead-extension cable system. If we suspect lead fracture, we must check the impedance of the electrode and to evaluate the side effects of voltage. Finally, we must conduct a radiological screening.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.