Abstract

A rapid, simple and sensitive liquid chromatographic method has been developed to assay ciclopirox olamine in raw material and topical solution. The analysis used a reversed-phase C18 end-capped column, with UV detection at 305 nm and pre-column derivatization with dimethyl sulphate. A linear response (r2 > 0.999) was observed in the range of 0.4–200 μg mL−1. The method showed good recoveries for the raw material and topical solution and the relative standard deviation intra and inter-day were ≤2.00%. Validation parameters such as specificity and robustness were also determined. The proposed method provided an accurate and precise analysis of ciclopirox olamine in raw material and topical solution.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.