Abstract

An LC-MS-MS method was revised and validated for simultaneous determination of icariin and its active metabolite icariside II in human plasma. The analytes and daidzein (IS) were extracted by liquid–liquid extraction and analyzed by LC-MS-MS. The separation was performed by a Zorbax SB-C18 column (3.5 μm, 2.1 × 100 mm) with an isocratic mobile phase consisting of methanol–water–formic acid (65:35:0.035, v/v/v) at a flow rate of 0.25 mL min−1. Detection was performed on a triple quadrupole tandem mass spectrum by multiple reaction monitoring mode using the electrospray ionization technique in positive mode. The method had lower limits of quantitation 0.2 and 0.1 ng mL−1 for icariin and icariside II, respectively, using 500 μL plasma sample. The linear calibration curves were obtained in the concentration range of 0.2–100 ng mL−1 for icariin and 0.1–100 ng mL−1 for icariside II. The RSD values of intra- and inter-day precision calculated from quality control (QC) samples were less than 7.2% for icariin and less than 6.5% for icariside II. The accuracy as determined from QC samples was within 3.8% for each analyte. The method has been applied to determine and evaluate the pharmacokinetic of icariin and its metabolite icariside II in volunteers following oral administration of icariin and extract of Epimedium, respectively.