Abstract

A sensitive and rapid LC–MS/MS method was developed and validated for the simultaneous quantitation of darolutamide and its active metabolite i.e. ORM-15341 in 50μL mice plasma using bicalutamide as an internal standard (I.S.) as per regulatory guidelines. Sample processing was accomplished through liquid-liquid extraction. Chromatographic separation was achieved using an Atlantis C18 column with an isocratic mobile phase comprising 0.2% formic acid:acetonitrile (35:65, v/v) at a flow rate of 0.8mL/min within 2.5min. Detection and quantitation were done by multiple reaction monitoring on a triple quadrupole mass spectrometer following the transitions: m/z 397→202, 395→202 and 429→255 for darolutamide, ORM-15341 and I.S, respectively in the negative ionization mode. The calibration curve was linear from 0.61–1097ng/mL for both darolutamide and ORM-15341. The intra- and inter-day precisions were in the range of 1.34–13.8 and 4.85–12.9 and 3.91–13.7 and 6.54–14.2%, for darolutamide and ORM-15341, respectively. Darolutamide and ORM-15341 were found to be stable under different stability conditions. The validated method was applied to a pharmacokinetic study in mice.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.