A reversed phase LC method was developed and validated to analyze the in vitro release of AZT from microemulsions. A mobile phase of acetonitrile:water (15:85) was used. The method validation showed good selectivity and linearity (r = 0.9993) for sample concentrations ranging from 0.6 to 100.0 μg mL−1. The RSD values (0.7–4.3%) and percentage recovery (88.1–109.8%) were within acceptable limits. The limit of detection (LOD) and limit of quantitation (LOQ) were found to be 0.012 and 0.041 μg mL−1. Quantitative analysis of the values obtained in the drug release assay indicates that the microemulsions used promote sustained release of AZT, which follows a Fickian diffusion mechanism.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.