Abstract

The layer thickness in rotational panoramic radiography is presented with the use of the concept of the noise-equivalent passband. Conventionally, the layer thickness has been calculated only from one-dimensional data in the horizontal or rotational dimension of panoramic radiography. In the present study, results from calculations using both one- and two-dimensional data are presented. When the vertical dimension is included in the calculations, the layer is wider than when calculated from data in the horizontal dimension only. It is pointed out that the wider layer that follows from the introduction of the vertical dimension may be the most reliable measure of layer thickness in panoramic radiography.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.