Abstract
Preliminary theoretical analyses indicate that lattice relaxation may be used to release lattice strain in the FAPbI3 perovskite to warrant both high x-ray detection performance and improved stability. Herein, we demonstrate stable black α-phase FAPbI3 single crystals (SCs) realized by lattice engineering via annealing in the ambient atmosphere. The engineered α-FAPbI3 SC detector shows almost all the best figures of merit including a high sensitivity of 4.15 × 105 μC Gyair-1 cm-2, a low detection limit of 1.1 nGyair s-1, a high resolution of 15.9 lp mm-1, and a short response time of 214 μs. We further demonstrate high-definition x-ray imaging at a dose rate below 10 nGyair s-1 on the FAPbI3 SC, indicating a minimal dose-area product of 0.048 mGyair cm2 to the patient for one-time posteroanterior chest diagnosis, which is more than 3000 times lower than the international reference level of 150 mGyair cm2. In addition, the robust long-term stability enables the FAPbI3 SC x-ray detector to work steadily for more than 40 years.
Published Version (Free)
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.